Four specialised air mattresses had interface pressure measured under six body sites prone to pressure sores in 10 subjects, supine and sitting. The mattresses were the Clinirest (SSI) and FirstStep (KCl) continuous airflow mattress overlays, and Airwave (Pegasus) and Nimbus (Huntleigh) alternating pressure air mattresses. On the mattress overlays, average supine interface pressures were 2.33 kPa (scapula), 4.15 kPa (elbow), 1.94 kPa (sacrum) and 2.79 kPa (buttock), although they were higher at the occiput (7.97 kPa) and heel (11.7 kPa). The alternating pressure air mattresses had an average minimum interface pressure close to zero for three sites, rising to 4.28 kPa under the heel. Average maximum interface pressures were 8.61 kPa (occiput), 5.21 kPa (scapula), 4.90 (elbow), 4.85 kPa (sacrum), 4.61 kPa (buttock) and 13.2 kPa (heel). No accepted scientific method exists for comparing the two types of mattress. Our data suggest a clinical benefit at the occiput and heel (supine) in using an alternating pressure air mattress and a benefit in using a continuous airflow mattress overlay at other sites.
Author(s): ALLEN V, RYAN DW, MURRAY A
Publication type: Article
Publication status: Published
Journal: BRITISH JOURNAL OF CLINICAL PRACTICE
Year: 1994
Volume: 48
Issue: 3
Pages: 125-129
Print publication date: 01/05/1994
ISSN (print):
ISSN (electronic):