ANGIOPLASTY PRESSURE-VOLUME MEASUREMENT

Lookup NU author(s): Emeritus Professor Alan MurrayORCiD

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

An instrument has been developed for controlled inflation and deflation of an angioplasty balloon. The mean difference between inflation to 800 kPa in air and a simulated coronary artery is 71.2 kPa. The repeatability of four inflation/deflation cycles is 6.4 kPa with the balloon in air and 9.0 kPa in the simulated artery.