Browse by author
Lookup NU author(s): Professor Tom Joyce
Full text for this publication is not currently held within this repository. Alternative links are provided below where available.
Although the market for replacement of diseased metacarpophalangeal (MCP) joints is dominated by single-piece silicone prostheses, several two-piece designs have been implanted. One such is the Digital Joint Operative Arthroplasty (DJOA) which consists of a part-spherical stainless steel metacarpal component which articulates within a matching concave phalangeal component made of ultra high molecular weight polyethylene (UHMWPE). A DJOA MCP prosthesis was tested using a clinically-validated finger simulator while a second DJOA prosthesis acted as a statically-loaded soak-control. Testing ran to 7.1 million cycles of flexion-extension. It was found that the UHMWPE components, both test and control, gained in weight by a similar amount. Therefore apparently there was no wear of the test components. However, the initial and final surface finish values of the test stainless steel metacarpal head were relatively high. Calculations based on this roughness data, plus recent dynamically-loaded soak data, may explain the apparent lack of wear.
Author(s): Joyce TJ
Publication type: Conference Proceedings (inc. Abstract)
Publication status: Published
Conference Name: 29th Northern Ireland Biomedical Engineering Conference
Year of Conference: 2010
Pages: 2337-2343
ISSN: 0957-4530
Publisher: Journal of Materials Science: Materials in Medicine, Springer
URL: http://dx.doi.org/10.1007/s10856-010-4010-z
DOI: 10.1007/s10856-010-4010-z
Library holdings: Search Newcastle University Library for this item
ISBN: 15734838