Toggle Main Menu Toggle Search

Open Access padlockePrints

Wear testing of a DJOA finger prosthesis in vitro

Lookup NU author(s): Professor Tom Joyce

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Although the market for replacement of diseased metacarpophalangeal (MCP) joints is dominated by single-piece silicone prostheses, several two-piece designs have been implanted. One such is the Digital Joint Operative Arthroplasty (DJOA) which consists of a part-spherical stainless steel metacarpal component which articulates within a matching concave phalangeal component made of ultra high molecular weight polyethylene (UHMWPE). A DJOA MCP prosthesis was tested using a clinically-validated finger simulator while a second DJOA prosthesis acted as a statically-loaded soak-control. Testing ran to 7.1 million cycles of flexion-extension. It was found that the UHMWPE components, both test and control, gained in weight by a similar amount. Therefore apparently there was no wear of the test components. However, the initial and final surface finish values of the test stainless steel metacarpal head were relatively high. Calculations based on this roughness data, plus recent dynamically-loaded soak data, may explain the apparent lack of wear.


Publication metadata

Author(s): Joyce TJ

Publication type: Conference Proceedings (inc. Abstract)

Publication status: Published

Conference Name: 29th Northern Ireland Biomedical Engineering Conference

Year of Conference: 2010

Pages: 2337-2343

ISSN: 0957-4530

Publisher: Journal of Materials Science: Materials in Medicine, Springer

URL: http://dx.doi.org/10.1007/s10856-010-4010-z

DOI: 10.1007/s10856-010-4010-z

Library holdings: Search Newcastle University Library for this item

ISBN: 15734838


Share