It has been reported that the introduction of a dielectric barrier between adjacent digits of an interdigitated electrode array can improve the sensitivity of the array as an electrochemical impedance biosensor. Here we present an in-depth analysis of the impedance in planar interdigitated electrodes and 3-D interdigitated electrodes (with dielectric barriers). The analysis indicates that the planar geometry not only provides lower impedance but also a higher change impedance as a result of molecular immobilization on the electrode array surface.