The Newcastle University research output collection, currently available on ePrints, will shortly be moving to a new open repository platform, Figshare. To prepare for the data migration we have paused adding new content to ePrints, and will resume once the new repository is launched. During this time you will continue to have access to ePrints (but no new content will appear). We will share updates here when available.

Transfer of low-molecular weight single-stranded DNA through the membrane of a high-flux dialyzer

Lookup NU author(s): Dr Nicholas Hoenich

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Purpose: Microbial contamination is often present in dialysate used for hemodialysis. Small single-stranded bacterial DNA sequences are capable of activating human inflammatory pathways, through mechanisms that include the Toll-like-receptor 9, and dialysis patients frequently show severe inflammation. Since these molecules have been found in dialysate and in patients' bloodstreams, we studied the potential of low-molecular weight DNA sequences, of the same structure as found in bacteria, to cross from the dialyzer circuit to the blood circuit of a dialysis filter.Methods: The mass transfer of DNA fragments across a high-flux dialyzer was evaluated with an in vitro dialysis model, in both conventional dialysis and pure convection mode. Measurement of DNA was performed by HPLC.Results: In dialysis mode, these mass transfer coefficients were calculated for different single-stranded DNA chain lengths: 5-bases = 28.5%, 9-bases = 20.5%, 20-bases = 9.4%, 35-bases = 2.4%, 50-bases and 100-bases, no transfer detected. In convection mode, these sieving coefficients were calculated: 5-bases = 1.0, 9-bases = 1.0, 20-bases = 0.68, 35-bases = 0.40, 50-bases = 0.17, 100-bases, no convective transfer detected. The physical size of DNA molecules could be the major factor that influences their movement through dialyzer pores.Conclusions: This study establishes that significant transfer across the dialyzer may occur with single-stranded DNA in the size range of 20-bases or less. These findings need to be confirmed with an in vitro whole blood model and with clinical investigations. Previous studies have described the clinical benefits of achieving high-purity dialysate. Precautions are warranted to minimize the presence of these DNA compounds in fluids utilized for hemodialysis treatment.


Publication metadata

Author(s): Tao X, Hoenich N, Handelman SK, Levin NW, Kotanko P, Handelman GJ

Publication type: Article

Publication status: Published

Journal: International Journal of Artificial Organs

Year: 2014

Volume: 37

Issue: 7

Pages: 529-538

Print publication date: 31/07/2014

Online publication date: 21/07/2014

Acceptance date: 30/04/2014

ISSN (print): 0391-3988

ISSN (electronic): 1724-6040

Publisher: Wichtig Publishing

URL: http://dx.doi.org/10.5301/ijao.5000338

DOI: 10.5301/ijao.5000338