Toggle Main Menu Toggle Search

Open Access padlockePrints

Evaluation of the HeartWare ventricular assist device Lavare cycle in a particle image velocimetry model and in clinical practice

Lookup NU author(s): Dr Stephan Schueler

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

OBJECTIVES: Ventricular blood stasis is a concern for continuous flow mechanical support devices and might contribute to the formation of thromboembolic events. The HeartWare (R) Ventricular Assist System (HVAD (R)) is equipped with the Lavare (TM) cycle that is a periodic speed modulation feature designed to alter flow patterns within the left ventricle and reduce areas of potential blood stasis. Here, we report in vitro and clinical findings on the effects of the Lavare cycle.METHODS: The effect of pump speed changes on the intraventricular flow field was examined with an in vitro particle image velocimetry model. The clinical impact of the Lavare cycle was evaluated through a retrospective review of the ReVOLVE study which includes 248 patients implanted with the HVAD following Conformite Europeenne Mark in nine centres in Europe and Australia. Baseline characteristics, adverse event profiles and Kaplan-Meier survival estimates were stratified by patients using/not using the Lavare cycle.RESULTS: Particle image velocimetry showed increased ventricular washout with an active Lavare cycle as measured by the fluid velocities and angular dispersion parameters. With the Lavare cycle on, there was also a 22% decrease in the stagnation index compared with when the Lavare cycle was off. In the ReVOLVE registry, patients with the Lavare cycle turned on (n = 215) were supported for 497 patient-years, whereas patients who did not use the speed modulation (n = 33) were supported for 39.3 patient-years. The Lavare cycle did not significantly affect patient survival as both groups had approximately an 80% survival after 1 year. Patients using the Lavare cycle had significantly fewer rates of stroke [0.06 vs 0.20 events per patient-year (EPPY), P = 0.0008], sepsis (0.03 vs 0.15 EPPY, P = 0.0003) and right heart failure (0.03 vs 0.18 EPPY, P < 0.0001) with no difference in the transplant or recovery rates among the two cohorts.CONCLUSIONS: The Lavare cycle effectively generates ventricular washout and the adverse event profiles of ReVOLVE patients with the Lavare cycle on were better than those with the Lavare cycle off. Larger studies are warranted to verify the positive effect of the Lavare cycle and to optimize speed modulation settings, so additional clinically relevant improvements can be realized.


Publication metadata

Author(s): Zimpfer D, Strueber M, Aigner P, Schmitto JD, Fiane AE, Larbalestier R, Tsui S, Jansz P, Simon A, Schueler S, Moscato F, Schima H

Publication type: Article

Publication status: Published

Journal: European Journal of Cardio-Thoracic Surgery

Year: 2016

Volume: 50

Issue: 5

Pages: 839-848

Print publication date: 01/11/2016

Online publication date: 07/09/2016

Acceptance date: 21/03/2016

ISSN (print): 1010-7940

ISSN (electronic): 1873-734X

Publisher: Oxford University Press

URL: http://dx.doi.org/10.1093/ejcts/ezw232

DOI: 10.1093/ejcts/ezw232


Altmetrics

Altmetrics provided by Altmetric