Study on a new electromagnetic flow measurement technology based on differential correlation detection

Lookup NU author(s): Professor Gui Yun TianORCiD

Downloads


Licence

This work is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).


Abstract

© 2020 by the authors. Licensee MDPI, Basel, Switzerland.Under the conditions of low flow rate and strong noise, the current electromagnetic flowmeter (EMF) cannot satisfy the requirement for measurement or separate the actual flow signal and interference signal accurately. Correlation detection technology can reduce the bandwidth and suppress noise effectively using the periodic transmission of signal and noise randomness. As for the problem that the current anti-interference technology cannot suppress noise effectively, the noise and interference of the electromagnetic flowmeter were analyzed in this paper, and a design of the electromagnetic flowmeter based on differential correlation detection was proposed. Then, in order to verify the feasibility of the electromagnetic flow measurement system based on differential correlation, an experimental platform for the comparison between standard flow and measured flow was established and a verification experiment was carried out under special conditions and with flow calibration measurements. Finally, the data obtained in the experiment were analyzed. The research result showed that an electromagnetic flowmeter based on differential correlation detection satisfies the need for measurement completely. The lower limit of the flow rate of the electromagnetic flowmeter based on the differential correlation principle could reach 0.084 m/s. Under strong external interferences, the electromagnetic flowmeter based on differential correlation had a fluctuation range in output value of only 10 mV. This shows that the electromagnetic flowmeter based on the differential correlation principle has unique advantages in measurements taken under the conditions of strong noise, slurry flow, and low flow rate.


Publication metadata

Author(s): Ge L, Chen J, Tian G, Zeng W, Huang Q, Hu Z

Publication type: Article

Publication status: Published

Journal: Sensors

Year: 2020

Volume: 20

Issue: 9

Online publication date: 28/04/2020

Acceptance date: 24/04/2020

Date deposited: 02/06/2020

ISSN (print): 1424-8220

ISSN (electronic): 1424-8220

Publisher: MDPI AG

URL: https://doi.org/10.3390/s20092489

DOI: 10.3390/s20092489

PubMed id: 32354018


Altmetrics

Altmetrics provided by Altmetric


Share