Browse by author
Lookup NU author(s): Professor Tom Joyce, Goksu KandemirORCiD, Andrew Bowey
This work is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).
© The Author(s) 2024.Purpose: Determine the performance of MAGEC X rods through retrieval analysis and comparison with clinical data. Methods: A multicentre explant database was searched to identify cases using MAGEC X device. Clinical and surgical data was gathered prospectively. Prior to rod disassembly, rods underwent testing with an external remote controller to measure the force output. Results: Eleven cases from 6 centres were identified. Implantation occurred at mean age 6.9 years with mean duration of implantation 42 months. Dual rod constructs were used in all but one case, providing 21 MAGEC X rods for explant analysis. Tissue metallosis was identified at revision surgery in 8/11 cases (73%). Of the 21 rods, 13/21 (62%) produced no force while 8/21 (38%) produced the force stated by the manufacturer. Endcap separation was seen in 5/21 (24%) rods. Where full disassembly was possible, 13/14 (93%) rods had an intact locking pin. Average rod growth was 18 mm (range 2–45 mm), equivalent to 5.1 mm per year. Wear debris was found within 20/21 (95%) MAGEC X rods. Conclusion: Despite the substantial design changes with MAGEC X tissue metallosis was seen in most cases; most explanted MAGEC X rods had lengthened only partially and produced no force output. While the previous issue of locking pin fracture appears to have been mitigated with MAGEC X, there are multiple other and new failure modes, such as endcap separation. Overall MAGEC X appears to give little improvement over earlier iterations of the rod.
Author(s): Joyce TJ, Kandemir G, Fender D, Bowey AJ, Rushton PRP
Publication type: Article
Publication status: Published
Journal: European Spine Journal
Year: 2024
Pages: epub ahead of print
Online publication date: 05/11/2024
Acceptance date: 22/10/2024
Date deposited: 19/11/2024
ISSN (print): 0940-6719
ISSN (electronic): 1432-0932
Publisher: Springer Science and Business Media Deutschland GmbH
URL: https://doi.org/10.1007/s00586-024-08546-6
DOI: 10.1007/s00586-024-08546-6
Altmetrics provided by Altmetric