Fully automated computer measurement of QT interval from the 12-lead electrocardiogram

Lookup NU author(s): Dr Philip Langley, Dr Fiona Smith, Dr Dingchang Zheng, Abigail Haigh, Emeritus Professor Alan MurrayORCiD

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

The aim was to develop a fully automatic QT interval measurement algorithm for the 2006 PhysioNet/Computers in Cardiology Challenge. The algorithm determined the Q onset and T offset points from the average beat in each lead. The QT interval measurement was calculated from the median Q onset and T offset points. Manual measurements were also made. The mean (sd) difference between automatic and reference measurements was -49 (38) ms and between manual and reference measurements was -6 (48) ms.


Publication metadata

Author(s): Langley P, Smith FE, King ST, Zheng D, Haigh AJ, Murray A

Editor(s): Murray, A

Publication type: Conference Proceedings (inc. Abstract)

Publication status: Published

Conference Name: 33rd Conference on Computers in Cardiology

Year of Conference: 2006

Pages: 345-348

ISSN: 0276−6547

Publisher: IEEE

URL: http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=4511859

Library holdings: Search Newcastle University Library for this item

ISBN: 9781424425327


Share