The Newcastle University research output collection, currently available on ePrints, will shortly be moving to a new open repository platform, Figshare. To prepare for the data migration we have paused adding new content to ePrints, and will resume once the new repository is launched. During this time you will continue to have access to ePrints (but no new content will appear). We will share updates here when available.

A focused-ion-beam-fabricated micro-paddle resonator for mass detection

Lookup NU author(s): Dr John HedleyORCiD

Downloads


Abstract

An experimental microelectromechanical system paddle resonator was made using focused ion beam fabrication from a 200 nm thick silicon nitride membrane and investigated using a laser vibrometer for resonant frequency detection. The fundamental mode was found to be torsional, as required, with a resonant frequency of 1.38 MHz and Q of 1070 at a test pressure of 20 μbar. The mass sensitivity was 55 ag Hz-1, close to the results of analytical and finite element analysis modeling, demonstrating proof of principle. © 2008 IOP Publishing Ltd.