Toggle Main Menu Toggle Search

Open Access padlockePrints

Examination of failed polyurethane gastrostomy devices and comparison with samples incubated in vitro

Lookup NU author(s): Professor Tom Joyce

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Percutaneous endoscopic gastrostomy (PEG) devices provide a route for nourishment and medication to those unable to swallow. The disintegration and discoloration of a type of polyurethane PEG device was evaluated. Five ex vivo samples were obtained and they showed discoloration, pitting and environmental stress cracking (ESC). A group of unused PEG devices were exposed to a gastric acid and pepsinogen formulation at 37 °C for up to 6 weeks. An additional cohort was investigated for fungal colonization by incubating with a Candida culture preparation at 37 °C for up to 6 weeks. Candida was shown to grow on all PEG samples exposed to it. The samples incubated in Candida hardened and discoloured and showed pitting but no ESC. In contrast, the samples incubated in the gastric acid formulation tended to soften and showed ESC but no discoloration. ©Società Italiana Biomateriali.


Publication metadata

Author(s): Joyce TJ

Publication type: Article

Publication status: Published

Journal: Journal of Applied Biomaterials and Biomechanics

Year: 2008

Volume: 6

Issue: 1

Pages: 30-34

Print publication date: 01/01/2008

ISSN (print): 1722-6899

ISSN (electronic): 1724-6024

Publisher: Wichtig Editore Srl


Share