The multicentre assessment of a procedure complexity index (CI) for the introduction of reference levels (RLs) in percutaneous transluminal coronary angioplasties (PTCA) is presented here. PTCAs were investigated based on methodology proposed by Bernardi et al. Multiple linear stepwise regression analysis, including clinical, anatomical and technical factors, was performed to obtain fluoroscopy time predictors. Based on these regression coefficients, a scoring system was defined and CI obtained. CI was used to classify dose values into three groups: low, medium and high complexity procedures, since there was good correlation (r = 0.41; P < 0.001) between dose-area product (DAP) and CI. CI groups were determined by an ANOVA test, and the resulting DAP and fluoroscopy time third quartiles suggested as preliminary RLs in PTCA, as a function of procedure complexity. PTCA preliminary RLs for DAP are 54, 76 and 127 Gy cm(2), and 12, 20 and 27 min for fluoroscopy time, for the three CI groups.
Author(s): Peterzol A, Quai E, Padovani R, Bernardi G, Kotre CJ, Dowling A
Publication type: Article
Publication status: Published
Journal: Radiation Protection Dosimetry
Year: 2006
Volume: 117
Issue: 1-3
Pages: 54-58
ISSN (print): 0144-8420
ISSN (electronic): 1742-3406
Publisher: Oxford University Press
URL: http://dx.doi.org/10.1093/rpd/nci719
DOI: 10.1093/rpd/nci719