Use of the profunda femoris to facilitate closure of an antegrade puncture with the StarClose device after proximal SFA angioplasty

Lookup NU author(s): Dr Sumaira MacDonald

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Purpose: To report a novel technique for safely closing antegrade common femoral artery (CFA) punctures using the StarClose device after proximal superficial femoral artery (SFA) angioplasty. Technique: The vessel locator of the StarClose device should not be deployed within a recently dilated vessel, so after proximal SFA angioplasty, the sheath is withdrawn into the CFA. A second guidewire is inserted into the profunda femoris artery followed by insertion of the StarClose sheath. The vessel locator is deployed in the profunda main stem and withdrawn into the CFA until resistance is felt, indicating apposition to the luminal aspect of the vessel wall. The device is subsequently deployed according to the manufacturer's instructions. Conclusion: Use of profunda femoris artery allows safe closure of the CFA using the StarClose device following antegrade puncture for proximal SFA angioplasty.


Publication metadata

Author(s): Yeung-Ngok-Kao S, Fisher RK, Williams RET, Jackson R, Rose J, Macdonald S

Publication type: Article

Publication status: Published

Journal: Journal of Endovascular Therapy

Year: 2006

Volume: 13

Issue: 4

Pages: 522-526

ISSN (print): 1526-6028

ISSN (electronic): 1545-1550

Publisher: Allen Press Inc.

URL: http://dx.doi.org/10.1583/06-1876R.1

DOI: 10.1583/06-1876R.1


Altmetrics

Altmetrics provided by Altmetric


Share