With greater numbers of primary knee replacements now performed in younger patients there is a demand for improved performance. Surface roughness of the femoral component has been proposed as a causative mechanism for premature prosthesis failure. Nineteen retrieved total knee replacements were analysed using a non-contacting profilometer to measure the femoral component surface roughness. The Hood technique was used to analyse the wear and surface damage of the matching ultra-high molecular weight polyethylene (UHMWPE) tibial components. All femoral components were shown to be up to 11x rougher after their time in vivo while 95 % showed a change in skewness, further indicating wear. This increase in roughness occurred relatively soon after implantation (within 1 year) and remained unchanged thereafter. Mostly, this roughness was more apparent on the lateral condyle than the medial. This increased femoral surface roughness likely led to damage of the UHMWPE tibial component and increased Hood scores.
Author(s): Scholes SC, Kennard E, Gangadharan R, Weir D, Holland J, Deehan D, Joyce TJ
Publication type: Article
Publication status: Published
Journal: Journal of Materials Science: Materials in Medicine
Year: 2013
Volume: 24
Issue: 2
Pages: 547-554
Print publication date: 01/02/2013
ISSN (print): 0957-4530
ISSN (electronic): 1573-4838
Publisher: Springer